7404-S |
7407-S |
7410-S |
Magnet pole cap face diameter |
50 mm (2.0 in) |
50 mm (2.0 in) |
50 mm (2.0 in) |
APPLIED FIELD STRENGTH (±1%) |
Room temperature
3.5 mm (0.14 in) sample access
|
2.63 T |
3.05 T |
3.42 T |
10 mm (0.4 in) sample access
|
2.27 T |
2.72 T |
3.16 T |
16 mm (0.64 in) sample access
|
1.94 T |
2.46 T |
2.92 T |
With oven or cryostat option
6.4 mm (0.25 in) sample access
|
1.49 T |
2.12 T |
2.59 T |
With single-stage variable temperature option
6.4 mm (0.25 in) sample access
|
1.75 T |
2.28 T |
2.76 T |
MOMENT MEASUREMENT |
Noise floor (emu RMS)
Room temperature
|
|
0.1 TC; 0.1 s/pt (no averaging) |
450 nemu |
0.1 TC; 1 s/pt |
175 nemu |
0.1 TC; 10 s/pt |
60 nemu |
With oven or cryostat option, 0.1 TC, 10 s/pt |
2.5 µemu |
With single stage variable temperature option 0.1 TC, 10 s/pt |
1.25 µemu |
With vector option 0.1 TC, 10 s/pt |
5 µemu |
Dynamic range |
1 × 10-7 to 103 emu |
Time constants (TC) |
0.1, 0.3, 1.0, 3.0, or 10.0 s |
Moment stability1 |
Less than 0.05% RMS of full scale/day at constant field, constant temperature, and highest range after a 5 h warm-up period |
Reproducibility |
Better than ±0.5%, or ±0.1% of full scale, fixed rotation angle and range, with sample replacement |
Moment accuracy |
Better than 1% of reading ±0.2% of full scale with a geometrically identical test sample and calibrant |
Sample mass |
0 to 10 g (higher mass can be accommodated with decreased performance) |
FIELD MEASUREMENT |
Field accuracy |
1% of reading or ±0.05% of full scale |
Field resolution |
2800 kA/m (35 kOe) |
|
8 A/m (0.1 Oe)
|
280 kA/m (3.5 kOe) |
0.8 A/m (0.01 Oe) |
28 kA/m (350 Oe) |
0.08 A/m (0.001 Oe) |
Closed loop field control stability |
<0.05% RMS of full scale/h |
MANUAL ROTATION |
Setting resolution |
<1° |
Setting reproducibility |
<1° |
Rotation range |
0 to 730° |